Capacity [Max] | 2,1g |
---|---|
Resolution [d] | 0,001mg |
Tare Range | – 6,1 g |
Repeatability (St. Dev.) (2 g) | 0,004 mg |
Repeatability (St. Dev.) (Max) | 0,005 mg |
Repeatability (St. Dev.) typical (2 g) | 0,004 mg |
Repeatability (St. Dev.) typical (Max) | 0,004 mg |
Minimum weight (USP) | 10 mg |
Minimum weight, typical (USP) | 8 mg |
Linearity Deviation, Typical | ± 0,01 mg |
Eccentric Deviation, Typical (Test Load) | 0,01 mg (2g) |
Response time (sec.) | ≤ 8 |
Calibration | Automatic internal |
Typical Sensitivity Offset | ± 2ppm/°C |
Display | 5″ Touch screen |
Operating languages | Italian, English, Portuguese, Deutsch, Francaise, Español, Chinese |
Interface | RS232, USB (import/export data) |
Pan size | 30mm |
Operating Environment | +15 °C – +30 °C , 45-65% Relative Humidity (non condensing) |
Operating temp. variation rate | ± 0,5 °C/h |
Relative humidity variation rate | ± 2%/h |
Verification | N/A |
Power Supply (balance) | 110-230Vac, 50/60Hz; output 24V 1A multi plug |
Power Supply (ioniser) | 110-230Vac, 50/60Hz; output 12V 1A multi plug |
Standard Accessories | Ioniser, E1 class 2 g weight, tweezer, gloves |
Balance Dimensions (WxLxH) | 165 x 480 x 205 mm |
Indicator dimension | 173(D) x 180(W) x 92(H) mm |
Distance pan/top wind shield | 75 mm |
Wind Shield internal diameter | 165 mm |
Repeatability is expressed as a standard deviation from 10 weighing cycles.
Stabilization time depends on the external environment conditions.